12 Operating Your Cleaner
Degassing
Step Action
1 Degas for 5-10 minutes. If necessary, use SET/CLEAR
DISPLAY to alter this setting. Default degas time is 5 minutes.
NOTE: Refer to page 17 for information on degassing.
2 Press ON/OFF once to start the degas process.
3 After completing the degas time, you are ready to set operating
parameters.
Setting Operating Parameters
Step Parameter Action
1 Set Time The cleaner is now in Set Time mode with a
default time of 60 mins. If necessary, use
SET/CLEAR DISPLAY to alter this setting.
Press ON/OFF once to activate timed
ultrasonics.
2 Set Temp. To set the tank temperature, press SELECT
OPTION until the SET TEMP LED lights. Then
press SET DISPLAY to alter the setting until the
LED display indicates the tank temperature you
wish to maintain. Press ON/OFF once to
activate heat. The heat indicator lights.
3 Solution
Temp.
To monitor the solution temperature, press
SELECT OPTION until the SOLUTION TEMP
LED lights. The LED display will indicate the
actual temperature of the solution.
NOTE:
If this is the first time you are running the cleaner, or if you have changed
cleaning solution, you must degas the solution. If not, move to Setting
Operating Parameters.